26
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3.
Using the 5 cm
3
(5 ml) syringe, slowly flush detergent solution through the
elevator wire channel via the washing tube several times until no air bubbles
exit the distal end (see Figure 3.14).
4.
Using the 5 cm
3
(5 ml) syringe, slowly flush water through the elevator wire
channel via the washing tube several times.
5.
Using the 5 cm
3
(5 ml) syringe, slowly flush air through the elevator wire
channel via the washing tube several times until a steady stream of air
bubbles exits the distal end (see Figure 3.14).
Figure 3.14
Disconnect the endoscope, reusable parts and cleaning
equipment
Do not touch the light guide of the endoscope connector
immediately after removing it from the light source because it
is extremely hot. Operator or patient injury may result.
1.
Disconnect the videoscope cable from the endoscope’s electrical connector.
2.
Disconnect the suction tube from the suction connector of the endoscope
connector.
3.
Disconnect the metal tip of the water container from the air/water supply
connector of the endoscope connector and attach the metal tip to the tip
receptacle on the lid of the water container as described in the water
container’s instruction manual.
4.
Disconnect the endoscope connector from the light source.
5.
Transport the endoscope to the reprocessing area.
5 cm3 (5 ml)
syringe
Elevator wire
channel
Endoscope
connector
Washing tube
Elevator channel
plug