Chapter 3 Cleaning, Disinfection and Sterilization Procedures
29
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3.4
Leakage testing
After precleaning, perform leakage testing on the endoscope to ensure that it is
waterproof. Also be sure to wear appropriate personal protective equipment.
Equipment needed
Prepare the following equipment:
Attaching the water-resistant cap (MH-553)
• The electrical connector of the endoscope is not waterproof.
Before immersing or leakage testing the endoscope, always
attach the water-resistant cap. Otherwise, equipment
damage may result.
• If the exterior of the electrical connector is scratched, the
connector may no longer be waterproof and the seal inside
the water-resistant cap may be scratched. If the electrical
connector is scratched, send it immediately to Olympus for
repairs.
• Always use a dry water-resistant cap. Any water remaining
on the water-resistant cap may cause damage to the
endoscope, maintenance unit or light source.
1.
Confirm that the inside of the water-resistant cap is dry and free from debris.
Wipe with a dry clean cloth if the inside of the water-resistant cap is wet or
there is debris.
2.
Align the OER/ETD/EW character (see Figure 3.17 (a)) or the EW character
on the water-resistant cap with mark 2 on the electrical connector housing.
When using a manual disinfector, align the KC/TD character (see Figure
3.17 (b)) on the water-resistant cap with mark 2 on the electrical connector
housing.
3.
Align the pin on the electrical connector with the groove on the
water-resistant cap.
• Personal protective equipment
• Large basin
• Clean water
• Maintenance unit or light source
(MU-1 or CLV-260, U40, U20)
• Leakage tester
(MB-155)
• Water-resistant cap
(MH-553)