Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
4.
Push the water-resistant cap into position and rotate it clockwise until it
stops (approximately 45°).
Figure 3.17
Performing the leakage test
• During leakage testing, a continuous series of air bubbles
emerging from a location on the endoscope indicates a leak
at that location. This means that water will be able to
penetrate the inside of the endoscope. If you locate a leak,
remove the endoscope from the water and contact Olympus.
• Never connect or disconnect the water-resistant cap or the
leakage tester’s connector cap while immersed. Doing so
could allow water to enter the endoscope and equipment
damage can result.
• Rotate the leakage tester’s connector cap until it stops. If it is
not fully and properly attached, the endoscope’s interior will
not be pressurized and accurate leakage testing will be
impossible.
Water-resistant
cap
(b) When using a manual disinfector
Rotate
Rotate
Mark 2 (gray)
2.
3.
4.
2.
3.
4.
Mark 2 (gray)
(a) When leakage testing, cleaning,
disinfection or using an
endoscope reprocessor