Chapter 3 Cleaning, Disinfection and Sterilization Procedures
47
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Presoak for excessive bleeding and/or delayed reprocessing
after each procedure
Follow the steps below only in cases of excessive bleeding
and/or delayed reprocessing; unnecessary immersions
should be avoided. Consecutive extended immersions may
damage the endoscope.
1.
Fill a basin with detergent solution at the temperature and concentration
recommended by the detergent manufacturer. Use a basin which is at least
40 cm by 40 cm (16” by 16”) in size and deep enough to allow the
endoscope to be completely immersed.
2.
Carefully coil the endoscope’s insertion tube and universal cord, and
completely immerse the endoscope in the detergent solution.
3.
Fill all channels with the detergent solution following the procedures
described in this section.
4.
Remove only the reprocessing equipment from the detergent solution.
5.
Cover the basin with a tight fitting lid to minimize the diffusion of detergent
vapors.
6.
Soak the endoscope for 10 hours at the temperature and concentration
recommended by the detergent manufacturer.
7.
Remove the endoscope from the detergent solution.
8.
After soaking the endoscope, manually clean it following the procedures
described in this section.