48
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3.6
High-level disinfection
All disinfection steps should be performed with the
endoscope and all equipment completely immersed in
disinfectant solution. If the equipment is connected to the
endoscope while they are immersed, or any part of them is
not immersed completely, disinfectant solution may not
adequately contact all surfaces of the equipment.
After manual cleaning, disinfect the endoscope according to the procedures
described below.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
Preparation
1.
Fill a basin with disinfectant solution at the temperature and concentration
recommended by the disinfectant manufacturer. Use a basin which is at
least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the
endoscope to be completely immersed.
2.
Attach the channel plug and the injection tube to the endoscope. Attach the
channel plug’s biopsy valve cap to the instrument channel port (see Figures
3.28 and 3.29).
3.
Connect the washing tube to the elevator channel plug (see Figure 3.31).
• Personal protective equipment
• Clean lint-free cloths
• Large basin
• Disinfectant solution
• 30 cm
3
(30 ml) syringe
• Channel plug
(MH-944)
• Injection tube
(MH-946)
• Washing tube
(MH-974)
• 5 cm
3
(5 ml) syringe