52
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Sterile water rinse
1.
Fill a basin with sterile water. Use a basin which is at least 40 cm by 40 cm
(16” by 16”) in size and deep enough to allow the endoscope to be
completely immersed.
2.
Immerse the endoscope, channel plug and injection tube in the sterile water.
Using a sterile lint-free cloth, thoroughly rinse and wipe all external surfaces.
3.
Immerse the washing tube in the sterile water. Using a sterile, lint-free cloth,
thoroughly rinse and wipe all external surfaces.
4.
Attach the channel plug and the injection tube to the endoscope. Place the
suction port in the sterile water. Attach the channel plug’s biopsy valve cap
to the instrument channel port.
5.
Using the 30 cm
3
(30 ml) syringe, flush the air/water and suction channels
with 90 cm
3
(90 ml) of sterile water respectively via the injection tube.
6.
Connect the washing tube to the elevator channel plug.
7.
Using the 5 cm
3
(5 ml) syringe, flush the elevator wire channel with 15 cm
3
(15 ml) of sterile water via the washing tube.
8.
Remove the endoscope, together with all equipment, from the sterile water
and place them in the sterile large basin.
9.
Using the 30 cm
3
(30 ml) syringe, flush 90 cm
3
(90 ml) of air through the
air/water and suction channels respectively via the injection tube.
10.
Using the 5 cm
3
(5 ml) syringe, flush the elevator wire channel with 15 cm
3
(15 ml) of air via the washing tube.
11.
Detach the injection tube only. Connect a sterile suction tube from the
suction pump to the suction connector on the endoscope. Turn the suction
pump ON and aspirate air for at least 15 seconds.
12.
Turn the suction pump OFF and disconnect all equipment from the
endoscope.
13.
Using a sterile lint-free cloth, thoroughly wipe and dry the external surfaces
of the endoscope and all equipment.
14.
Dry the endoscope and all equipment.
15.
Store the components following the instructions given in Chapter 5, “Storage
and Disposal”.