Chapter 3 Cleaning, Disinfection and Sterilization Procedures
57
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Manual cleaning
• Make sure that the items immersed in detergent solution do
not contact one another.
• Make sure not to scratch the seals on the air/water valve and
AW channel cleaning adapter with brushes, etc.
1.
Fill a small basin with clean water. Use a basin which is deep enough to
allow all equipment to be completely immersed.
2.
Fill a small basin with the detergent solution at the temperature and
concentration recommended by the detergent solution. Use a basin which is
deep enough to allow all equipment to be completely immersed.
3.
Detach the biopsy valve’s cap from the main body, before immersing in the
detergent solution (see Figure 3.35).
Figure 3.35
4.
Immerse all other equipment in the detergent solution.
5.
Using a soft brush or clean lint-free cloth, clean the external surfaces of all
equipment in detergent solution.
6.
Using a soft brush or clean lint-free cloth, clean the inside of the mouthpiece
in detergent solution.
7.
Using the channel cleaning brush, thoroughly brush the openings of the
suction valve, air/water valve and AW channel cleaning adapter until no
debris can been seen.
8.
Using the channel cleaning brush and channel-opening cleaning brush,
thoroughly brush the interior and openings of the biopsy valve.
9.
Clean the bristle of the cleaning brushes thoroughly while the brushes are
immersed.
Main body
Cap