CV-190/190PLUS/290+EU-ME2 type E1S2/2
95
[SETTING INFORMATION]
Setting
model
Step
Default
Setting
parameter
CV-1500
Setting > System setup > Video output > 12G-SDI OUT
(4K/HD) signal
HD:1080p
4K
Setting > System setup > Peripherals 1 > Keyboard type
MAJ-1923
US MAJ-1710/
1995 (EN)
Setting > System setup > Peripherals 2 > Select PIP/POP
device
CV
CV
Setting > System setup > Peripherals 2 > Ultrasound center
type
Not connected
EU-ME1/
EU-ME2
Setting > System setup > Peripherals 2
>CV PIP/POP US input terminal
HD/SD input
(SDI)
HD/SD input
(SDI)
Setting > System setup > Peripherals 2
> US/multi-monitor Fixed EVIS output
Disabled
Enabled
Setting > System setup > Peripherals 2 > 12G-SDI OUT
(4K/HD)
Fixed EVIS
Fixed EVIS
Setting > System setup > Peripherals 2 > 3G-SDI OUT (HD)
EVIS/US
switch
EVIS/US switch
Settings > User Presets > New entry or Edit > Release/Pre-
freeze
> Record ext. source
ON
OFF
Setting > Users Presets > New entry or edit > PIP/POP >
Display method
PIP
PIP
Setting > Users Presets > New entry or edit > PIP/POP
> PIP position
Lower left
Lower left
Setting > Users Presets > New entry or edit > PIP/POP
> Select CV ext. source
HD/SD input
(SDI)
HD/SD input
(SDI)
EU-ME2
System setup > Video signal > Video signal format >
MONITOR OUT
HDTV(RGB)
HDTV(RGB)
System setup> Video signal > HD aspect ratio > MONITOR
OUT
4:3
4:3
System setup > Remote > Remote > EVIS
Video system
Center
Video system
Center
System setup > Remote
>Monitor, keyboard setting (CV combination)
1 M
1 Keyboard
1 M
1 Keyboard
OEV321UH
MENU > Preset > Load Preset
B/C
B/C
MENU > Input/Output Configuration > Port A Input Select
12G-SDI 1
12G-SDI 1
OEV262H
Preset > Load preset
B/C
B/C
PORT A
SDI
HD15
* Please refer to CV-190/190PLUS/290 for the setting of connection with CV-190/190PLUS/290.
Summary of Contents for EVIS X1
Page 1: ...EVIS X1 Installation Guide Ver 4 0 Olympus Europa SE Co KG 1...
Page 4: ...4 1 System organization...
Page 6: ...6 2 Package contents...
Page 8: ...8 3 Precautions for installation...
Page 15: ...15 4 Set up device on WM NP3...
Page 17: ...17 5 Installation and setting...
Page 24: ...24 6 CV 1500 wiring and setting Peripheral equipment excl US center...
Page 43: ...43 7 CV 1500 wiring and setting US Center...
Page 69: ...69 8 CV 1500 wiring and setting CV 190 190PLUS 290 And peripheral equipment...
Page 108: ...108 9 CV 1500 wiring and setting ENDO AID CADe...
Page 109: ...109 Cabling Standard Setup ENDO AID CV 1500 OEV321UH...
Page 112: ...Alternative with Scopeguide OIP 1 CV 1500 UPD 3 112...
Page 113: ...Setup 3a OIP 1 CV 1500 UPD 3 IMH 200 113...
Page 114: ...Setup 3b OIP 1 CV 1500 UPD 3 IMH 10 20 114...
Page 115: ...Case 4 OIP 1 CV 1500 UPD 3 nCare 115...
Page 116: ...Case 5 OIP 1 CV 1500 UPD 3 nCare EUS 116...
Page 117: ...Case 6 CV 1500 OEV321UH OIP 1 PSCU 117...
Page 118: ...Case 7 CV 1500 OEV321UH OIP 1 PSCU nCare 118...
Page 119: ...Case 8 CV 1500 OEV321UH OIP 1 PSCU HVO 400ST 119...
Page 120: ...Case 9 CV 1500 OEV321UH OIP 1 UPD 3 PSCU 120...
Page 121: ...Case 10 CV 1500 OEV321UH OIP 1 UPD 3 PSCU Care 121...
Page 122: ...Case 11 CV 1500 OEV321UH OIP 1 UPD 3 PSCU HVO400 122...
Page 123: ...123 10 Appendix...
Page 128: ...128...