background image

Olympus FV1000                                                                         -6-                                                                  Short instructions 
 

4. Using the climatic chamber 

1.

 

Check the water bottle at the control unit behind the microscope table. Refill with 
distilled water only. Distilled water is in Wolfgang Posselt’s office.  

Eventually open the valve underneath the water bottle.

 

2.

 

Turn on the power switch at the right side of the control panel no.

 . 

3.

 

Switch on the microscope following the procedure in chapter 3. The climatisation 
control software opens automatically. 

4.

 

In the software set the desired temperature between 30° and 42° C. Cooling is not 
possible. 

Do not activate the humidity before the temperature is stable to avoid 

condensation which will damage the microscope equipment.

 

5.

 

If you are using the CO

2

 open the gas bottle at the wall behind the microscope table. 

Turn the main valve counter clockwise, not more than a quarter turn. 

6.

 

When the temperature is stable set the desired values for CO

2

 concentration (between 

0 and 10%) and relative humidity.  

Don’t set the relative humidity higher than 60%. Higher values will cause 
condensation which will damage the microscope equipment. The software will 
open a warning. Confirm by pressing no. 

7.

 

For shutting down, select first the dry mode in the software. The values for CO

2

 and 

humidity will be set to minimum and the temperature to 37°C. 

Let the dry mode run 

until the humidity is less than 25%.

  

8.

 

Close the gas bottle, if used.. 

9.

 

Turn off the power switch at the control panel no. 

.  

2

Summary of Contents for Fluoview FV1000

Page 1: ...the microscope system For a detailed instruction in system operation refer to the Olympus user manual This instruction also contains important information on laser safety and general guidelines Conten...

Page 2: ...r eyesight may be permanently damaged A reflected laser beam is just as dangerous as a direct beam 3 Never deactivate the laser protection devices Read the Safety Guide in the Olympus User s Manual to...

Page 3: ...hese hours pick up a key in advance When you are leaving lock the microscope room and after 17 00 also look the main door of the facility Return the key to Wolfgang Posselt or to the mail box of the L...

Page 4: ...7 Lasers Lasers for the lines 491 532 and 559 nm have to be switched on manually Read chapter 3 for details 8 Data transfer Never store any data on the microscope computer There are three ways to tra...

Page 5: ...nm are controlled entirely through the Olympus software These lasers are switched on by selecting an acquisition protocol from the dye list in the Image Acquisition window or by directly clicking on...

Page 6: ...able to avoid condensation which will damage the microscope equipment 5 If you are using the CO2 open the gas bottle at the wall behind the microscope table Turn the main valve counter clockwise not m...

Page 7: ...microscope 1 Exit the control software 2 Shut down Windows 3 Turn the key switch of the 491 532 nm laser to off no 4 4 Turn the key switch of the 559 nm laser to off no 5 5 Switch off all buttons of...

Reviews: