APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 7 1
Page
2-2-4 Image Acquisition by Rotating It (Rotation Scan)
Rotation scan enables image acquisition with tilting the field of viewed.
Rotation Scan can not be used with Free Line Scan.
1. Acquire an image in the XY observation mode. For the operating procedure, see
section 2-2-1, “Image Acquisition in XY Observation Mode”.
2.
Set the image acquisition mode.
For details, see sections 2-2-1, “Image Acquisition in XY Observation Mode”, 2-2-2,
“Image Acquisition in Other Observation Modes”, 2-2-5, “Image Acquisition of Only the
Rectangular Position (Clip Scan)”, 2-2-6, “Image Acquisition by Magnifying the
Rectangular Position (Zoom-In Scan)”, 2-2-7, “High-Speed Image Acquisition” and 2-
2-8, “Acquiring Finer Image (Sequential Scan)”.
3. Specify the image rotation angle using the rotation arrow buttons on both sides of the
circuit in the [Pan] group box in the [Acquire] panel.
4.
Acquire the image by clicking the <XYZ>, <XYT> or <XYZT> button.
NOTE
Dragging the blue handle
along the circuit allows the
acquisition area to be tilted.
Assuming that the angle at
the top of the circle is 0°, tilts
the acquisition area by 1° per
step in the counterclockwise
direction.
Assuming that the angle at
the top of the circle is 0°, tilts
the acquisition area by 1° per
step in the counterclockwise
direction.
Displays the image rotation
angle. The angle can also be
specified by direct entry.