APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 8 3
Page
3. In the [Mode] group box in the [Acquire] panel, click the [Surface XY-Norm] option
button, select <Sequen> from the list below it and select the scan mode to be used from
the displayed icons.
Do not select <Line Aq> in the [Mode] group box in the [Acquire] panel
because the virtual channels are not compatible with Line Sequential
Scan.
If the observation mode is set to Sequential Scan, the [Dyes] sub-panel in the
[Acquire] panel cannot be used.
4.
Using the <
> and <
> buttons in [Group], set the group according to the reagent in
use. This operation is not required if the current setting needs not be changed.
(Example: CH1 (FITC) and CH2 (PI) are set to Group 1.)
NOTE
TIP
Click to increase the acquired group No.
Click to decrease the acquired group No.
Shows the group to be acquired.