APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 8 6
Page
Setting will disappear if <Normal> is selected at [Mode] group box of
[Acquire] panel after setting virtual channel. It is useful that the setting is
saved with use of <Save> button at [Acquisition Settings] group box on
[Settings] sub panel of [Acquire] panel before selecting <Normal>.
It is not permitted to set the excitation dichroic mirror differently for each
channel. The excitation dichroic mirror set up at the last is applied to all
groups.
When the laser to be used with a group has not been set, the [Laser Unit]
dialog box shows “No Laser has assigned yet”.
In this case, set the laser in the [Laser Unit] group box.
If Sequential Scan is set already, the <C.A. Adjustment> button in the [Optical
System Configuration] window cannot be selected.
7.
Adjust the image brightness for every group.
Select the group to adjust the image brightness in the [Group] drop-down list displayed
on the upper part of the [Acquire] panel.
Select a group before repeated scanning. The group selection cannot be
changed during repeated scanning.
NOTE
TIP
TIP
[Group] drop-down list
Select the sequential scan
group.
TIP
NOTE