APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 1 0 8
Page
3.
From the page tabs on the bottom right of the [Acquire] panel, select the [Time Series]
sub-panel. The panel as shown below appears.
Fig. 2-27 [Time Series] sub-panel
4.
Set the interval time to “10” using the <
> and <
> buttons in the [Interval] text box.
Change the interval time corresponding to the specimen.
5.
Set the number of scans to “30” using the <
> and <
> buttons in the [N] text box in
the [Time Series] sub-panel. Change the number of scans corresponding to the
specimen.
6. Confirm that the time for image acquisition is set to 5 minutes (or 300 seconds) in the
[Total Time] text box. The time for image acquisition differs according to the interval
and the number of scans.
6 Setting the TIME COURSE Software (optional)
Setting of the real time graph is required when you use the TIME COURSE software.
See the “FV-TIEMPO TIME COURSE software User’s manual” for details of the TIME
COURSE software.
[Rest Time] text box
Shows the time after the end of an
image acquisition until the start of next
acquisition
[N] text box
Set the number of scans with the <
>
and <
> buttons. Key entry is also
acceptable.
Under text box, it displays max. number
of slices that can be acquired with use
of physical memory only available at
that time.
[Interval] text box
Set the interval with the <
> and<
>
buttons. Key entry is also acceptable.
[Total Time] text box
Shows the time required to complete image
acquisition
.