APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 1 1 1
Page
6.
Start the third scan in 10 seconds after the second scan is performed.
(The [Disabled] option button is selected.)
7.
Start the forth and subsequent scan in 10 seconds after the third scan is performed.
The image of the photobleached specimen is acquired.
When performing the forth or subsequent scan, be sure to switch the [REX
mode] and [Bleach mode] between scans.
In the TIME COURSE software (optional), when the REX mask file in the
REX or Bleach mode is changed while image scanning is performed
obtaining the real time graph, the laser is irradiated while the pop-up menu
is displayed, however, the real time graph is not plotted. Please be aware
this before using.
The image of the specimen
is fading only where the
laser is irradiated.