APPLIED OPERATIONS
/Image Processing
OPERATION INSTRUCTIONS
2 - 2 3 4
Page
2-6-2 Contrast Conversion
The LUT intensity can be mapped (re-assigned) while observing a histogram.
Mapping (re-assignment) results in changing the image contrast.
An image acquired by observation contains intensity information in values from 0 to 4095,
but the intensity information used in actual display takes values from 0 to 255 by assigning
the original values from 0 to 4095 to values from 0 to 255 usually. This facility changes the
contrast by noticing a certain section between 0 and
4095 and mapping this section to values between 0
and 255.
1.
Display the [Process] panel at the front.
2.
Display the [Histogram] sub-panel of the [Process] panel at the front.
Fig. 2-87 [Histogram] Sub-panel
3.
Display the image window of the image you want to change the contrast.
The icon of the image is displayed in the frame at the top left of the [Process] panel
0
4095
mapping
0
255
0
255
Sets the image to be subjected to contrast
change. The icon of the image is
displayed here.
<Compute Histogram> button
Displays the histogram.
[Channel] option button
Select the channel to be subjected to
LUT intensity mapping (re-assignment).