Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
OPERATION INSTRUCTIONS
1 - 1 2
Page
1-2-2 Focusing on the Specimen
1-2-2-1 Combination with BX
1.
Select the light path for 100% eyepiece by pushing in the light path selector
knob (1) fully to the stop position.
2. From the page tabs on the bottom right of the [Acquire]
panel
, select the
[Settings] sub-panel.
Fig. 1-3
[Settings] Sub-panel
3. Select the <BI> button in the [Light Path] group box.
The <BI> button looks pushed in to indicate that it is selected.
4. Push the cube button (5) of the hand switch (4) to engage the optimum
cube for specimen dye. In the cube display window (2), the selected cube
is displayed.
5. When you do the transmitted light observation, push the transmitted light
DIC slider (3) and let it into the light path.
6. Focus on the specimen by looking into the eyepiece. Be sure to adjust
the diopter of the eyepiece in advance. (Refer to the instruction manual of
the BX microscope.)
(2)
(4)
(5)
(1)
(3)