Software Setup/
Setting the System Configuration
MAINTENANCE
1 - 1 5
Page
(Combination with IX)
When using IX, three parts can be motorized among the filter turrets, the
mirror unit, and condenser turret.
[Mirror Unit] group box
Sets up the name and color of the cube
turret.
[Nosepiece] group box
Sets up the name, color of the objective
and sets the magnification, N.A,
refractive index (n), number of the
confocal pinhole, and the condenser
worked with to each objective.
[Condenser Turret] group box
Sets up the name and color of the
universal condenser.
[Filter Turret] group box
Sets up the name and color of the filter
turret for reflected light observation.
[Filter Turret] group box
Sets up the name and color of the filter
turret for visual observation.
[Filter Turret] group box
Sets up the name and color of the filter
turret for transmitted observation.
NOTE
[Shutter] group box
Sets up to use the shutter for
transmitted observation.
[Shutter] group box
Sets up to use the shutter for reflected
light observation.