There’s no general rule: these adjustments mainly depend on your sample.

 

4.6 Image Adjustments 

 Setting the Gain and Offset properly 

Check image saturation level by clicking 
Hi-Lo icon. The image will 
change color to show red and 
blue pixels. Blue indicates that the signal 
in a given pixel is black, with a value of 0. 
Red indicates that the signal in a given 
pixel is saturated (over-exposed) with a 
value of 4095. 

For optimal performance, the image 
needs to be so that then entire image is 
set to the limits of the dynamic range. 
Different sets of experiments may have 
different fluorescence intensities. 

Adjust the image using a combination of 
the excitation 

Laser power

HV 

and 

Offset

Reduce either the HV or the laser power, 
until red pixels disappear. Increase Offset until a dusting of blue pixels in the background are 
observed. 

Click 

LSM start 

to capture an image.  

4.7 How to do a Z stack for a 3D Image (xyz) 

1.

 

Under Z Series: 

ON

 for Z. 

2.

 

Select 

Live 2x 

or

 Live 4x

 to see the 

specimen and click 

Register

 

under origin 

tab to save your focal plane where your 
image is focused. After completing the Z-
stack. 

Move

 

can be selected to move back 

to the focal plane of focus. 

3.

 

Turn the focus knob clockwise and move all 
the way to the top of your sample. Click 

Register

 under start tab. 

4.

 

Turn the focus knob anticlockwise and move 
all the way to the bottom of your sample. 
Click 

Register

 under end tab. 

5.

 

With 

system optimized

 

selected, the 

program will determine the optimal settings 
for your Z-stack. You can override these 
settings using the 

slices

 or 

z-step size

 boxes. 

Beware that changing these will alter the z-
resolution you attain. 

 

 

 

 

 

 

Summary of Contents for FV3000

Page 1: ...oking and System Access 4 3 1 Acknowledgements 5 4 Operation Procedures 5 4 1 Switching ON Protocol 5 4 2 Objective lens 6 4 2 Software Initiation 7 4 3 Setting up directory to save images 8 4 4 Laser...

Page 2: ...Immersion Oil Modern immersion oil has no known hazards to human beings so far yet they can cause discomfort to a person if the immersion oil has been left on skin for too long or inhaled Therefore pl...

Page 3: ...taff before they are allowed to access the confocal system 2 1 To arrange a training session Please email Bioimaging Facility bioimaging tll org sg for a training session Fill in the necessary particu...

Page 4: ...er 6pm Sat Sun Public holidays If users require more slots in a particular week they can book on weekdays during non peak hours after 6pm on weekends Sat and Sun and on public holidays If extra time i...

Page 5: ...RT TIMING b Switch on the Master switch on the wall 1 PC 2 Wait until windows icon pops up c Switch on Stage controller box 3 CBH controller 4 d Optional Turn on the Metal Halide lamp 5 if needed e Tu...

Page 6: ...t to 10x objective or lower then click YES Cleaning the stage will take some time b Otherwise click NO to skip cleaning 4 2 Objective lens Objective lens Immersion media 1 25x 0 04 Plan Apochromat Air...

Page 7: ...n the stage mechanism This adds time to the initialization process Do not place anything on the stage mechanism until the initialization process is complete If you are not the first user of the week C...

Page 8: ...y to save all your images Please save in your folder You have several different options on where you can save your images Option 1 D or E Users shaalini create your own folder BPAE cells oir Do keep i...

Page 9: ...oduce intense background noise SD keep HV within 600 800 More than 800 will contribute to intense background noise Under PMT settings tab Mode Select VBF Average average a given number of scans to get...

Page 10: ...to zoom images only in the center Ideal zoom which satisfy the Nyquist limit can be achieved by zooming till the red arrow any zoom beyond it will result in poor resolution and result in more photoble...

Page 11: ...the laser power until red pixels disappear Increase Offset until a dusting of blue pixels in the background are observed Click LSM start to capture an image 4 7 How to do a Z stack for a 3D Image xyz...

Page 12: ...your timelapse experiment Interval FreeRun continues to capture images until it has completed your selected cycle You could always choose an interval of your choice 4 9 Reuse previous settings and ex...

Page 13: ...h the sample and return the stage to central position d Carefully clean the objective lens with Whatman lens tissue using 100 ethanol e Switch lens back to 10x 0 3 objective lens f Exit the FV31S SW s...

Page 14: ...ission please contact Bioimaging Group No stage movement Check the coarse fine movement setting near the focus and xy position knobs It is a physical button that has no electronic indication elsewhere...

Page 15: ...id white or coloured light sometimes red light might be indicative of error III Restart the main controller right at the bottom if none of the steps taken work especially if the machine has been left...

Page 16: ...and cannot be recovered Restart computer terminal Startup popup error code LAST_OPERATION_ NONCOMPLETED The last user did not shut down the program properly or the program encountered a fatal error du...