Laser intensity change
/
Laser emission area and Laser intensity change for AOTF combiner (using FV300 or FV500)
IV.
OPERATION INSTRUCTIONS
IV.
3 - 2
Page
3-2 Laser emission area and Laser intensity change for AOTF combiner
(using FV300 or FV500)
When FV300 or FV500 is used and laser combiner is arranged with AOTF (FV5-
COMBA), laser intensity and laser emission area (REX) can be changed. For further
details, see instruction manual for FV300 or FV500. Laser intensity can be changed
from 1% to 100% if the maximum value is set to 100%.
1.
Verify that FV5-COMBA is connected with FV5-LCU.
2.
Connect FV5-LCU with FV5-LDPSU.
3.
Set REX from FV300 or FV500 software.
4.
After REX setting, start scanning so that laser emission area can be changed.
As in the case of FV5-COMB(2), laser intensity can also be changed with
scale setting of [Laser Intensity] group box. In case that FV5-COMBA is
used, the laser intensity can be changed from 1% to 100% if the maximum
value is set to 100%.
Set back panel switch of FV5-LDPSU properly. Regarding DIP switch
setting, refer to VI – 1 [Back panel switch settings].
When you use this system as standalone, refer to Chapter 2 [Laser
emission].
NOTE
TIP
TIP
Summary of Contents for FV5-LD405
Page 2: ......
Page 8: ......
Page 10: ......
Page 20: ......
Page 22: ......
Page 40: ......
Page 42: ......
Page 44: ......
Page 46: ......
Page 48: ......
Page 50: ......
Page 52: ......
Page 54: ......
Page 62: ......
Page 64: ......
Page 66: ......
Page 74: ......
Page 76: ......
Page 78: ......
Page 80: ......
Page 82: ......
Page 84: ......
Page 86: ......
Page 90: ......
Page 91: ......