Scanner Assembly and Setup
35
Before you install the new transducer on the wedge, ensure that there is a sufficient
amount of couplant between the transducer and the wedge.
10. Tighten the transducer on the wedge.
Do not overtighten the transducer on the wedge; overtightening might crack the
wedge.
11. Reinstall the transducer assembly on the bearing arm.
2.2.6
Replacing a Phased Array Probe
To replace a phased array ultrasonic probe, the probe assembly must be removed and
partially disassembled.
To replace a phased array probe
1.
Turn off the couplant flow.
2.
Disconnect the couplant tubes from the couplant source.
3.
Disconnect the probe cable on the instrument.
To prevent any damage to the Hypertronics connector pins, be sure to always install
the protective cover when the probe cable is disconnected.
4.
Remove the scanner from the inspection surface.
5.
Using a flat screwdriver, at the back of the bearing arm, unscrew the screw to free
the yoke from the bearing arm (see Figure 2-26 on page 33).
6.
If it is necessary to remove the yoke from the wedge, use a hexagonal key to
remove the screw (on the side of the yoke) that holds the yoke parts together (see
Figure 2-29 on page 36).
NOTE
CAUTION
CAUTION
Summary of Contents for GLIDER
Page 6: ...DMTA029 01EN Rev E December 2017 List of Abbreviations vi...
Page 16: ...DMTA029 01EN Rev E December 2017 Important Information Please Read Before Use 10...
Page 18: ...DMTA029 01EN Rev E December 2017 Introduction 12...
Page 54: ...DMTA029 01EN Rev E December 2017 Chapter 2 48...
Page 62: ...DMTA029 01EN Rev E December 2017 Chapter 4 56...
Page 70: ...DMTA029 01EN Rev E December 2017 Chapter 5 64...
Page 76: ...DMTA029 01EN Rev E December 2017 List of Tables 70...