DMTA029-01EN, Rev. E, December 2017
Specifications
53
4. Specifications
The general specifications for the GLIDER scanner are listed in Table 3 on page 53.
Table 3 General specifications
Parameter
Value
General
Dimensions and weight
Motion increment
4 mm
Mounting pod holding
force — direction axial to
mounting post (under
ideal conditions)
Manual suction cups (76 mm [3 in.] OD): 7 kg (15 lb) per cup
Manual suction cups (114 mm [4.5 in.] OD): 18 kg (40 lb) per cup
Venturi suction cups (102 mm [4 in.] OD): see Table 4 on page 54
Magnetic pods: 81 kg (180 lb) per pod
Encoder
Type: Quadrature
Resolution: 13 steps/mm ± 0.15 steps/mm (330 steps/in. ±
0.006 steps/in.)
Waterproof
Pinout: see Figure 4-1 on page 55
Power rating: 5 VDC, 25 mA maximum
Frequency: 0 kHz to 1.5 kHz (for a maximum displacement
velocity of 10 cm/s [4 in./s])
Environment
Outdoor use
Yes
Altitude
Up to 2000 m
Operating temperature
5 °C to 40 °C
Storage temperature
−
20 °C to 60 °C
Relative humidity (RH)
Max. 80 %, non condensing