23
4 Pre-observation/pre-operation preparation and inspection
4.3
Pre-operation/post-operation inspections
Inspection location
Item
If abnormality is
discovered
Main unit
Any dirt, sand or other foreign object caught in the
battery cover, video connector cap or AC adapter
connector cap?
See "7 Storage
and maintenance"
(P. 44).
Battery cover cracked or scratched?
Replacement
required.
Contact Olympus.
AC adapter connector cap or video connector cap
torn or chipped?
Irregularities such as cuts or kinks in the universal
cable?
Control unit
Irregularities such as damage to or deformation of
the exterior material, buttons or levers?
Fastening ring on the control unit loose?
Continued use in this condition creates the risk of
the angulation control unit falling off during use.
Insertion tube,
angulation sec-
tion, distal end
Any deformation, looseness or unusual wear on the
distal end, angulation section, or insertion tube?
Never use the instrument while the distal end is
loose. Doing so creates the risk of loose parts fall-
ing off during use.
See "7 Storage
and maintenance"
(P. 44).
Any dirt or water droplets on the objective lens of the
distal end?
Continued use in this condition creates the risk of
the distal end becoming damaged.
Insertion tube broken or crushed?
If the insertion tube is deformed, it might no longer
be able to be inserted into or drawn out of the
observation object.
LCD monitor
Any surface cracks or other abnormalities?
NOTE
The LCD monitor is manufactured based on preci-
sion technologies. The LCD monitor may contain pix-
els that do not light (visible as black dots) or light
permanently (visible as bright dots). This does not
indicate a defect or malfunction of the product.
Replacement
required.
Contact Olympus.
Any fingerprints on the screen or other dirt?
See "7 Storage
and maintenance"
(P. 44).
Carrying case
Any damage, looseness or other irregularities on the
exterior, handle, latch, or other parts of the carrying
case?
Summary of Contents for IPLEX TX
Page 1: ...INSTRUCTIONS INDUSTRIAL VIDEOSCOPE IPLEX TX ...
Page 3: ......