14
}The power supply unit to be used is variable depending on the illumination column in use.
· IX2-ILL100: TH4 power supply unit.
· IX2-ILL30: TL4 power supply unit.
USING THE CONTROLS
Fig. 4
@
²
³
|
4-1 Power Supply Unit and Microscope Frame
1
(Figs. 4 & 5)
1. Make sure that the light intensity control knob @ is in the MIN (minimum
intensity) position and set the main switch ² to “
I
” (ON). (The POWER
LED ³ lights up.)
2. Rotate the knob @ toward MAX (maximum intensity) to increase the
intensity and the illumination brightness.
}With the TH4, the position (approx. 9 V) marked indicates the position
where the daylight illumination suitable for photomicrography is obtained
when the 45LBD filter is engaged in the light path.
The position marked of the IX71 and TH4-HS also has the identical
function.
With the TL4, the position marked is not provided and even
the maximum light intensity (6 V) may sometimes be unsuitable
for photomicrography.
With the IX71 Frame (Fig. 6)
}If the TH4 connection cable provided with the microscope frame is
connected to the TH4 (the REMOTE LED | lights in this case), the light
intensity control knob @ on the power supply unit is defeated and only
the light intensity control knob ƒ on the front of the microscope frame
is available.
1. Press the transmitted light ON-OFF button … (so that the button is illumi-
nated) and adjust the brightness with the light intensity control knob ƒ.
2. To turn the lamp OFF, set the transmitted light ON-OFF button … to OFF.
# The microscope frame is in standby mode when the REMOTE LED
| is lit. Power of about 2.5 W is consumed in this period.
When the microscope system is not be used for a long period, set
the main switch ² to “ ” (OFF).
Turning Power On, Adjusting the
Brightness
Fig. 5
@
²
Fig. 6
ƒ
…