MIC-D D.ADJUSTMENT PROCEDURE
D-1
Tension gauge
Tension gauge
frame
shaft
1.Rotating force adjustment of ARM and HEAD
(1)Tools
-1-Tension gauge
(2)Adjustment
-1-Measure the rotating force of arm and head by tension gauge.
-2-Check that the amount of each rotating force are in a standard.
[STANDARD] ARM : 4.5-6N
HEAD: 1-2N
-3-In besides a standard, adjust by bolting of frame or shaft.
2.Rotating force adjustment of Focusing Knob
(1)Tools
-1-Tension gauge
(2)Adjustment
-1-Measure the rotating force of focusing knob by tension gauge.
-2-Check that the amount of rotating force is in a standard.
[STANDARD] :0.5-1N
(3)In besides a standard, adjust by bolting of focusing knob or shaft.