MIC-D D.ADJUSTMENT PROCEDURE
D-4
‑9‑Fix the C‑MOS
Y(CUKSK3X6SA:4pcs.) and Lightintensity adjustment board
(CUKSK3X6SA:2pcs.) by screws.
‑10‑Remove the scale seal, lose the screws.(ACU3X4SA:2pcs.)
scale seal
ACU3X4SA(2pcs)
‑11‑Set up the cross specimen on stage.
‑12‑Conect the USB cable with PC.
‑13‑Set up the zoom to highest magnification position.
‑14‑Focus the specimen by knob.
‑15‑Set up the zoom to lowest magnification position.
‑16‑Control the parfocality lens by MIC‑DKC02 knob.
‑17‑Apply the adhesive(SE9176) to screws.(ACU3X4SA:2pcs.)
A C U 3 X 4 S A
Adjustment lens
Knob