}Use the fluorescence mirror unit that is optimum for each purpose of observation.
4-3 Coaxial Fluorescence Illuminator
1
Selecting the Fluorescence Mirror Unit
Mirror unit name
Dichroic mirror
Excitation filter
Barrier filter
Application
U-MCFPHQ/XL
DM450HQ
BP425-445HQ
BA460-510HQ
ECFP
U-MGFPHQ/XL
DM485HQ
BP460-480HQ
BA495-540HQ
EGFP
U-MYFPHQ/XL
DM505HQ
BP490-500HQ
BA515-560HQ
EYFP
U-MRFPHQ/XL
DM565HQ
BP535-555HQ
BA570-625HQ
RFP
U-MGFP/XL
DM505
BP460-490
BA510IF
EGFP
U-MGFPA/XL
BA510-550
Spectral Characteristics of the Filters
1. Filter for fluorescent protein CFP U-MCFPHQ/XL
2. Filter for fluorescent protein GFP U-MGFPHQ/XL
T
ransmittance %
Wavelength nm
3. Filter for fluorescent protein YFP U-MYFPHQ/XL
4. Filter for fluorescent protein RFP U-MRFPHQ/XL
U-MGFPA/XL :
For dye separation
T
ransmittance %
Wavelength nm
T
ransmittance %
Wavelength nm
T
ransmittance %
Wavelength nm