![background image](http://html1.mh-extra.com/html/olympus/mvx10/mvx10_instructions-manual_4110997046.webp)
42
10
(Figs. 42 to 44)
}For the lamp housing for xenon burner, refer to the instruction manual
provided with the lamp housing. The centering method is identical to
that for the mercury burner.
Attaching the Mercury Burner
1. Using the Allen screwdriver, loosen the socket clamping screw @.
2. Hold the upper section of the lamp housing and pull it upward to remove
the socket section.
# To prevent malfunction, do not hold the lamp housing by the
centering knobs ².
3. Place the socket section upside down as shown in Fig. 43.
}The lamp housing is equipped with the holder for transportation in the
factory shipment condition or with an old burner after the burner is
replaced. Remove the holder or old burner by loosening the two burner
holding screws ³.
4. Attach the + (positive) pole of a specified mercury burner | to the fixed
mount on the upper side, then the - (negative) pole to the mount on the
lower side.
# Be sure to use the USH-103OL (OLYMPUS) or the HBO103W/2
(OSRAM) mercury burner.
Be careful and avoid leaving fingerprints or contaminants on
the mercury burner. Otherwise, there is a danger of explosion
due to distortion of glass caused by the stains. If the burner is
contaminated, clean it by wiping gently with gauze slightly
moistened with absolute alcohol.
5. Attach the socket section with burner to the original position and tighten
the socket clamping screw @.
# Align the external edges of the lamp housing with those on the
socket section, and push the lamp housing straight downward.
Fig. 42
@
²
Fig. 43
³
|
Mounting the Lamp Housing for
Mercury Burner
Burner Service Life
USH-103OL: 300 hours
}This value assumes light cycles composed of 2 hours of lighting
and 30 minutes of extinction. Do not turn it on and off at a shorter
cycle than the above, for this will shorten the service life of the burner.
After replacing the burner, reset the hour counter to “0.0” as
outlined above.
Summary of Contents for MVX10
Page 2: ......
Page 13: ...9 MVX10 Make a photocopy of this spread and post it near the microscope for quick reference...
Page 52: ...MEMO...
Page 53: ...MEMO...
Page 54: ...MEMO...
Page 55: ......