DMTA-10040-01EN, Rev. E, February 2018
Chapter 4
114
4.4.4
Filter Menu in Dual Frequency — MAIN FILTER Key
HI PASS
filters
High-pass filters may be set from 0 Hz (
OFF
) to 100 Hz in 1 Hz increments, and
from 100 Hz to 1000 Hz in 5 Hz increments. To adjust the
HI PASS
filter settings,
press the MAIN FILTER menu key (
) three times, followed by the A key, and
then rotate the knob to the desired value.
LO PASS
filters
Low-pass filters may be set from 10 Hz to 100 Hz in 1 Hz increments, up to
500 Hz in 5 Hz increments, and up to 2000 Hz in 25 Hz increments, followed by
wide band. To adjust the
LO PASS
filter settings, press the MAIN FILTER menu
key (
) three times, followed by the B key, and then rotate the knob to the
desired value.
CONT NUL
(continuous null)
CONT NUL
allows a very low-frequency high-pass filter to be turned on, which
is useful to keep the null point of the eddy current probe at the specified point, if
required. When turned on, this function adds a 0.2 Hz, 0.5 Hz, or 1 Hz high-pass
filter. By default, this feature is set to
OFF
.
To turn on continuous null, press the MAIN FILTER menu key (
) three times,
followed by the C key, and then rotate the knob to the desired value.
SCAN RPM
(NORTEC 600S and NORTEC 600D models only)
The
SCAN RPM
function controls the scanner rotational speed (revolutions per
minute) when an optional rotating scanner is connected to the NORTEC 600S
(N600S) or NORTEC 600D (N600D) instrument models.
To adjust the
SCAN RPM
setting, press the MAIN FILTER menu key (
) three
times, followed by the E key, and then rotate the knob to the desired value.
4.4.5
Special Menu in Dual Frequency — MAIN FILTER Key
PRB DRV
(probe drive)
The NORTEC 600 instrument has three levels of probe drive that can be selected:
LOW
,
MEDIUM
, and
HIGH
. The approximate peak-to-peak voltages are 2 V, 6 V,
and 12 V, respectively.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...