DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
145
probe’s longest face still parallel to the notches), and then press the A-LIFT NULL
key (
) to scan the entire standard.
The resulting signal is shown in Figure 5-36 on page 145.
Figure 5
‑
36 The signal after scanning the entire standard
To fine-tune the instrument settings
1.
Depending on your requirements, set the alarm parameters, horn, or external
horn (louder). For more details about alarms, see “Alarm Menus” on page 292.
2.
Depending on your requirements, set the display erase or persistence values to
automatically refresh the screen. For more details about screen erase options, see
“
PERSIST (variable persistence)
page 89.
You can also check if any other grids and screen display modes might be useful
for your purposes; for example,
SWP+IMP
. For more details about screen options,
see “Display Menu — DISP Key” on page 87 and “Display Menu in Dual
Frequency — DISP Key” on page 116.
3.
Confirm that the value of the maximum signal amplitude and signal angle are
displayed by default, as shown in the example in Figure 5-37 on page 146. For
more details about the reading type or position in the impedance plane display,
see “Displaying Real-Time Readings” on page 56.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...