Using the Instrument
165
Figure 5
‑
60 The Alarm parameters
2.
Depending on your requirements, set the display erase or persistence values to
automatically refresh the screen.
For more details about screen erase options, see “
PERSIST (variable persistence)
3.
Press the FULL NEXT key (
) to toggle to the full-screen mode, and then scan
the standard.
The resulting signal should resemble the example shown in Figure 5-61 on
page 166. Note that the value of the maximum signal amplitude and the signal
angle are displayed by default. For more details about the reading type or
position in the impedance plane display, see “Displaying Real-Time Readings” on
page 56.
The list of all parameters is shown in Figure 5-62 on page 166.