DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
169
6.
Press the ANGLE key (
), and then adjust the lift-off (probe-motion) angle so
that it is as horizontal as possible and the notch signal points upward (see
Figure 5-65 on page 169).
Figure 5
‑
65 Adjusting the signal angle
7.
Press the GAIN key (
), and then decrease the horizontal gain until the notch
signal is contained within the first horizontal division (10 % horizontal) [see
Figure 5-66 on page 169].
Figure 5
‑
66 Adjusting the gain