DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
213
28. While holding the probe on position 1, insert a shim (one business card) under the
probe.
29. While firmly pressing the probe on the business card, adjust the
ANGLE
(B key)
until the mix signal (orange) reaches the same vertical height as the null point (see
Figure 5-131 on page 213).
Figure 5
‑
131 Adjusting the ANGLE value
30. Press the MAIN FILTER menu key (
) once.
31. While firmly pressing the probe on the business card, adjust the
H GAIN 2
(D key) until the mix signal extends about 5 horizontal divisions away from the
null point (see Figure 5-132 on page 213).
Figure 5
‑
132 Adjusting the H GAIN 2 value