Figure 5
‑
167 The frequency 2 scan signal of the support ring
6.
Slowly scan the 1.3 mm (0.052 in.) thru-wall hole only, pressing the ERASE foot
switch as required to clear the screen. When the hole signal is visible on the
NORTEC 600 screen, press the FREEZE key (
) [see Figure 5-168 on page 240].
Figure 5
‑
168 The frequency 2 scan signal of the 1.3 mm (0.052 in.) thru
‑
wall hole
7.
Press the ANGLE key (
), and then rotate the frequency 2 signal until the
hole signal reaches a phase of approximately 40°.