DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
274
Figure 5
‑
213 The scan of the two grooves
3.
Press the ANGLE key (
), and then rotate the signal until the OD grooves are
split by the vertical cross-hair (see Figure 5-214 on page 274).
Figure 5
‑
214 Adjusting the signal phase
4.
Press the GAIN key (
), and then decrease the gain until the 60 % wall-loss
flaw signal reaches approximately 90 % of vertical height.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...