Figure 5
‑
238 The 40 % wall
‑
loss groove (
left
) and 60 % wall
‑
loss groove (
right
)
6.
Press the DISP menu key (
), followed by
DSP MODE
(display mode, A key),
and then rotate the knob until
SWP + IMP
(sweep plus impedance) is displayed.
7.
Press
SWP TIME
(D key) and adjust the sweep time to accommodate the length
of the tube being inspected.
8.
Place the probe in a defect-free area of the calibration standard near the 60 % wall-
loss flaw, and then press the NULL foot switch.
9.
Slowly scan the tube.
The scan result should be similar to that shown in Figure 5-239 on page 292. The
SWP
(sweep) display on the left of the screen clearly shows the flaws in the tube.
The larger indications (full screen) represent the 40 % and 60 % wall-loss flaws.
The center (very small) indication represents the thru-wall hole.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...