DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
217
•
Tube probe adaptor (differential) including NULL and ERASE foot switch (P/N:
N600-TUBE-FTSW [U8779875]) and auxiliary absolute tube adaptor (P/N: DGL-
AF4-BNC-8 [U8779886])
•
Admiralty brass calibration standard (tube): OD 19 mm (0.75 in.), WT (wall
thickness) 1.65 mm (0.065 in.); P/N: CT02-001-D16 [U8779241]
To set the initial NORTEC 600 configuration
1.
Connect the probe and the foot-switch adaptor cable to the PROBE connector on
the NORTEC 600.
2.
Press the ADV SETUP menu key (
) once, then select
APPL SELECT
(A key)
to open the application selection menu. Select
Heat Exchanger Tubing
with the
knob, then press
to accept (see Figure 5-136 on page 217).
Figure 5
‑
136 The Heat Exchanger Tubing application
3.
Press the MAIN FILTER menu key (
), and then set the
FREQ
(A key) to
14 kHz.
4.
Press the MAIN FILTER menu key (
) again, and then set the
FREQ 2
(A key)
to 3 kHz.
To calibrate the signals
1.
Place the probe in a defect-free zone of the calibration standard, and then press
the NULL foot switch.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...