Chapter 2 Compatible Reprocessing Methods and Chemical Agents
OSF-3 ENDOSCOPE REPROCESSING MANUAL
• Inspect the instrument package for openings, tears and other
damage. If the instrument package is open or damaged, seal
the components in new packaging and re-sterilize as
described below.
• Allow the packages to dry within the autoclave, using the
autoclave’s drying cycle (if available) or by opening the door
of the autoclave and allowing the packages to air dry.
Handling a wet package can compromise its sterility.
Endo-Therapy Accessories and other items which are
marked by the words “AUTOCLAVE” or “AUTOCLAVABLE”,
a green band or yellow-green marking are compatible with
autoclaving.
Temperature
Exposure Time
Prevacuum
132 to 134
°
C
(270 to 274
°
F)
5 minutes
Table 2.4