Chapter 3 Cleaning, Disinfection and Sterilization Procedures
25
OSF-3 ENDOSCOPE REPROCESSING MANUAL
4.
Install the Leakage Tester Connector cap on the Venting Connector of the
Endoscope Connector (see Figure 3.11).
Figure 3.11
5.
With the Leakage Tester connected, immerse the endoscope in water and
observe for approximately 30 seconds while angulating the Bending Section
of the endoscope. Confirm that there is no location on the endoscope from
which a continuous series of bubbles emerges.
6.
Remove the endoscope from the basin.
7.
Turn the Maintenance Unit or the Light Source OFF.
8.
Disconnect the Leakage Tester from the Maintenance Unit or Light Source.
9.
Wait 30 seconds, or until the covering of the Bending Section contracts to its
pre-expansion size. Disconnect the Leakage Tester connector cap from the
venting connector.
10.
Thoroughly dry the Leakage Tester.
Connecting the Leakage Tester
(MB-155)
MU-1
Light Source
2.
4.