Chapter 3 Cleaning, Disinfection and Sterilization Procedures
OSF-3 ENDOSCOPE REPROCESSING MANUAL
Figure 3.16
3.
Immerse the Filter of the Intake Tube in the detergent solution.
4.
Attach the 30 cm
3
(30 ml) syringe to the Main Body’s Luer-lock.
5.
Using the 30 cm
3
(30 ml) syringe, inject 150 cm
3
(150 ml) of detergent
solution into the Air/Water and Suction Channel to flush the channels.
6.
Disconnect the All-channel Irrigator from the Endoscope, and leave all items
immersed.
The Luer lock of the syringe will provide a secure connection.
However, do not use excessive force when connecting it, as
the Luer lock on the Main Body may be damaged. In addition,
do not hold the Intake Tube’s port when operating the
syringe, as this part may be damaged.
Air/Water Cylinder
Lock cover
Suction Cylinder
Channel cap
Instrument
Channel Port