Chapter 3 Cleaning, Disinfection and Sterilization Procedures
33
OSF-3 ENDOSCOPE REPROCESSING MANUAL
Figure 3.17
Soaking all Components in Detergent Solution
1.
Using a clean, lint-free cloth, wipe all debris from the endoscope’s outer
surface while the endoscope remains immersed in the detergent solution.
2.
Cover the basin with a tight fitting lid.
3.
Soak the endoscope and All-channel Irrigator (CW-3), for the amount of
time and at the temperature recommended by the detergent manufacturer.
Removing and Rinsing the Endoscope and all Equipment
Remove the Endoscope and all equipment from the detergent solution and place
in clean water.
Syringe
Intake tube (with filter)
Luer lock
Main body
Lock cover
Air/Water Cylinder
Suction Cylinder
Universal Cord
Channel cap
Air Channel
Water Channel
Air Nozzle
Water Nozzle
Insertion Tube
Water Container
Connector
Suction
Connector
Air feeding and water feeding conditions
of the Air and Water Channels from
Section 3.3, “Precleaning” to Section 3.7,
“Rinsing After High Level Disinfection”.