Chapter 6 Care, Storage and Disposal
133
VIDEO SYSTEM OTV-SI
Chapter 6
Care, Storage and Disposal
6.1
Care
•
Use only dry gauze to clean the power switch. Do not
moisten the gauze, as the switch could fail or cause an
electric shock.
•
After cleaning this instrument, dry it thoroughly before use. If
it is used while wet, there is a risk of an electric shock.
•
When cleaning the video system, wear personal protective
equipment to guard against dangerous chemicals and
potentially infectious material, such as eye wear, a face
mask, moisture-resistant clothing and chemical-resistant
gloves that fit properly and are long enough so that your skin
is not exposed. Blood, mucus and other protenially infection
control risk.
•
Do not clean the video connector socket and the output
connector on the front panel, or the terminals or AC power
inlet on the rear panel. Touching them can deform or corrode
the contacts and cause equipment damage.
•
Do not autoclave or gas sterilize this instrument. Equipment
damage will occur.
•
Do not wipe this instrument’s surface with an abrasive
material. The instrument’s surface will be scratched.
If the equipment is soiled, perform the following cleaning procedures
immediately after use. If cleaning is delayed, organic debris will solidify and it
may be difficult to effectively clean this instrument. The equipment should also
be cleaned routinely.
1.
Turn this instrument OFF and disconnect the power cord from the wall
mains outlet.
2.
Should the equipment become soiled with blood or other potentially
infectious materials, first wipe off all gross debris using a neutral detergent,
then wipe with a lint-free cloth moistened with a surface disinfectant.
VISY218
Summary of Contents for OTV-SI
Page 2: ...VISY218...
Page 6: ...Contents iv VIDEO SYSTEM OTV SI VISY218...
Page 148: ...142 Chapter 7 Troubleshooting VIDEO SYSTEM OTV SI VISY218...
Page 156: ...VISY218...
Page 159: ...INSTRUCTIONS COMPACT TROLLEY TC C2 VISY218...
Page 171: ...VISY218...
Page 172: ...VISY218...
Page 173: ...VISY218...