Using Software Options
Part # 910-250C
97
Figure 50
DAC Curve
Note:
If at any time the operator needs to delete a point during DAC setup, the DEL
PT key [F4] deletes the last point taken for the DAC curve. If the operator
needs to delete ALL points taken for a particular DAC setup, the DEL ALL
key [F3] deletes all points taken during setup. These functions are accessible
only during DAC curve setup.
Once the Custom DAC Curve settings are defined and the DAC points have been captured
and completed, the operator can toggle between DAC and TVG views, manipulate
RANGE, ZERO OFFSET and ANGLE, and add necessary scanning gain, curve gain
adjustment, or transfer correction. The TVG view of any Custom DAC Curve includes the
user-defined reference curves as well as the primary DAC curve. Custom DAC Curve also
incorporates the reference correction functionality, if desired.
Figure 51
Custom DAC Curve