Using Software Options
Part # 910-250C
105
be used to store a reference waveform for an Undersized Weld and [2
nd
F], [F3] can be
used to store a reference waveform for a Stick Weld, and so on. A reference waveform can
be updated by simply saving over the current saved reference waveform by pressing [2
nd
F], [F1] again.
The reference waveforms are stored in the battery-backed memory of the EPOCH 4PLUS.
This means they are saved even when the unit is powered off. These reference waveforms
are kept in these memory locations until new waveforms are saved in the [F1] – [F5] keys
or until you recall a file with a different set of saved overlays.
9.7.3
Accessing Datalogger Storage
Datalogger File Locations 1-9 (Storage of Waveform Overlays)
Datalogger file locations 1-9 are also known as “Quick Recall Calibration Files.” You can
save ALL data associated with a Spotweld Inspection Setup in these files. Pressing the
[SAVE WAVE] key when the Spotweld Overlay option is active stores the following
items in file locations 1-9:
•
Live waveform (or frozen waveform if the [FREEZE] key is used)
•
Setup parameters
•
All reference waveform overlays assigned to the [F1] – [F5] keys
Note:
The overlays stored with the setup are NOT visible in the datalogger screen
or from the Interface Program. However, the overlays are stored with the
setup and will be recalled to their respective [F1] – [F5] key locations when
recalled to the live screen. Press the [CALIBRATION] key followed by the
number of the desired file in the datalogger [1] – [9] to complete this recall
procedure.
Datalogger File Locations 10 and higher (Storage of Weld Rating Stamping)
Datalogger file locations 10 and higher are generally used to save data acquired during the
actual inspection of Spotweld parts. When saving data to files 10+, you can stamp the
individual Identifier (ID) with a two-letter code representing the quality of the weld by
pressing the [SAVE THICK] or [SAVE WAVE] key.
•
[SAVE THICK] saves the ID with only the thickness reading.
•
[SAVE WAVE] saves the ID with the thickness reading, waveform, and setup
parameters.
After one of these keys is pressed, the following prompt appears:
SELECT STAMP TYPE TO SAVE
F1 = GOOD
F2 = UNDERSIZED
F3 = STICK
F4 = WELD A
F5 = WELD B