Using Software Options
Part # 910-250C
147
Figure 99
AWS D1.1 Software Option Activated in the Options Menu
When AWS D1.1 is activated, the function keys on the EPOCH 4PLUS provide the
following options:
[F1]: REF B – Store the dB value from a reference reflector.
[F2]: No function.
[F3]: No function.
[F4]: SCAN DB – Toggle the scanning gain between a selected value and zero dB.
[F5]: AUTO-80 – Adjust a gated signal to 80% FSH.
After the AWS option is activated from the Options Menu, you must set a REF B value in
order to begin an inspection. This number represents the gain level necessary to bring the
echo from a reference reflector to 80% Full Screen Height (FSH). The reference reflector
is often a Side Drilled Hole in the calibration block used for the angle beam calibration.
Other reference reflectors may be used provided that they meet AWS requirements for
these inspections.
In order to store a REF B value, you must gate the echo from the reference reflector and
bring the echo to 80% FSH. You can achieve this by pressing the [GAIN] key and
adjusting the gain with the Up and Down Arrow keys, or by pressing [F5] to automatically
adjust the gain to bring the echo to 80% FSH. Once the echo has the appropriate
amplitude, press [F1]. The following screen appears: