This Quick Reference Guide provides instructions to use the Cell Counter model R1. For detailed instructions, please refer to the user manual in the supplied USB memory. 
  

Using the R1 with Trypan Blue Stain 

1.  Select protocol 

  Select 

protocol

 from the main menu. 

  Select the DEFAULT or desired protocol. 

  Make sure the dilution factor is set to 2 to count with 

trypan blue.  

  Press 

Load

 to apply the selected protocol.  

 

2.  Prepare & load sample 

  Mix 10 µL cell suspension with 10 µL trypan blue. 

Pipette gently.  

  Open a new Cell Counting Slide.  

  Hold the Cell Counting Slide by its edges and load 

10 µL of the cell sample into a sample chamber. 

 

3.  Insert slide 

  Insert the Cell Counting Slide face up and sample-

side first into the counting slide port of R1. 

 

!

 

Important!

  Do not insert the Cell  Counting Slide 

facedown.  

4.  Count cells 

  Select 

count

 from the main menu.  

  Make sure the staining option is “With TB”.  

  Press the circle next to 

[Autofocused Counting]

The circle will turn blue when the autofocus is 

activated.

 

 

  Press 

[Count]

.

 

 

 

Results

 

1.  Save and print results 

  Insert a USB memory.  

  Press 

[Save/Print] 

in the Results screen. 

  Select the desired saving options (analyzed image, raw image, report). 

  Name the count with the onscreen keyboard. Press 

Add Date/Time

  to add the date 

and time to the name.  

  Press 

Save

. A folder will be created to contain the generated files.  

  Press 

Print

 to get a printed report of the cell count results. 

 

2.  Image View 

  Press 

[Image]

 to view the captured image of the analyzed cell sample.  

  Press the magnifier button to zoom in and out of the image. 

  Press 

Tag

  to label what was counted as live cells with green circles and dead cells 

with red circles.  

 

3.  Histogram and Gating 

  Press 

[Histogram & Gating]

  to see a graphical representation of the cell count 

results. 

  Press 

Total/

on

. The button will become 

Live/

on

  and display the size distribution of 

only live cells. Press 

Live/

on

. The button will become 

Dead/

on

  and display the size 

distribution of only dead cells. 

  Select a light grey limit icon. Press the arrows to alter the size limits.  

  Press 

Apply

 to set the size gating limits. The count results will adjust accordingly.  

  Press 

Cell number

 to change the Y-axis to 

Cell Concentration

.  

  Press 

Cluster Map/off

  to change it to 

Cluster Map/on

  and show the distribution of 

cell clusters.  

 

4.  Dilution Calculations 

  Press 

[Dilution]

 to view the dilution calculator.  

  Press 

[Total]

  to change the current concentration to 

[Live]

, the concentration of live 

cells in the cell suspension.  

  Input the values into the blanks for the desired final concentration and volume.  

  Press 

Calculate

 to receive directions for dilution.  

 

 