} Weight of each unit: 51.8 kg (STM7-SF), 53.8 kg (STM7-SFA), 77.1 kg
(STM7-MF), 78.6 kg (STM7-MFA)
} Weight applied on one transport stick:
Front transport stick
Back transport stick
of the frame
of the frame
STM7-SF
Approx. 1 kg
Approx. 25 kg
STM7-SFA
Approx. 1 kg
Approx. 26 kg
STM7-MF
Approx. 10.1 kg
Approx. 28.4 kg
STM7-MFA
Approx. 10.3 kg
Approx. 29 kg
4. Carry the frame STM7-LF/STM7-LFA using the transport sticks with
both hands.
} Weight of each unit: 150.5 kg (STM7-LF), 152 kg (STM7-LFA)
} Weight applied on one transport stick:
Front transport stick
Back transport stick
of the frame
of the frame
STM7-LF
Approx. 33.6 kg
Approx. 41.6 kg
STM7-LFA
Approx. 34 kg
Approx. 42 kg
·
To prevent malfunction, hold the stage only by the transport
rod or the transport sticks.
·
Be sure to detach the stage, observation tube, reflected
light arm and digital indicator before carrying the frame. If
you hold parts other than the frame, the product may be
damaged.
5. Carry the stage STM7-CS50/STM7-CS100 using the transport rods
with both hands.
} Weight of each unit: 18.1 kg (STM7-CS50), 17.7 kg (STM7-CS100)
} Weight applied on one transport rod: Approx. 9 kg (STM7-CS50),
Approx. 8.9 kg (STM7-CS100)
6. Carry the stage STM7-CS200/STM7-CS300 using the transport sticks.
} Weight of each unit: 59.8 kg (STM7-CS200), 111.3 kg (STM7-CS300)
} Weight applied on one transport stick: Approx. 15 kg (STM7-CS200),
Approx. 27.8 kg (STM7-CS300)
7. Be sure to check the equipment before use.
Particularly, prior to using the stage, check the stage by moving it
slowly making one reciprocation back and forth and from side to
side over the entire movable area.
Doing so will spread the grease at the guide area over the entire stage
to keep the stage operations in a good condition for a long time.
8. Do not touch the connector pins of modules with your hands
directly.
The modules may be damaged by static.
9. To perform the measurement as accurately as possible, warm up
the equipment for about one hour (Z-axis manual type) or about two
hours (Z-axis motorized type) after turning ON the powers of all units
until starting the measurement.
10. With the Z-axis motorized type model, be sure to set the lower
software limit in order to prevent collision between the objective and
measuring object (page 32).
11. With the Z-axis manual type model, do not rotate fine focusing
knobs by holding both right and left fine focusing knobs a as
shown in the left picture. Otherwise, the fine focusing knob may
come off. If it comes off, contact Olympus.
CAUTION
CAUTION
a
STM7-CS50/STM7-CS100
STM7-CS200/STM7-CS300
STM7-LF/STM7-LFA