72
Bundle the cables loosely giving plenty of length so that they
won't be stretched when operable parts move.
Particularly, do not bundle the cables connected to the stage
or the focusing unit with tension in all movement ranges.
} The cable distribution is illustrated in combination with the large
motorized frame as a reference. Unless otherwise specified, the
same applies to other combinations.
} Attach the cable holder to the position illustrated in the picture and
distribute the cables.
CAUTION
16
Cable distributions
Cable holder a
Cable holder b
Fit the cables in the cable holder a
, giving the cable a certain slack so that
it does not hinder the vertical movement
of the Z-axis.
Secure the cables in the cable holder b giving
the cable a certain slack so that it does not
hinder the vertical movement of the Z-axis.
Cable holder b for Digital Indicator (STM7-DI)
While the stage is moved toward the front
right, insert the cables to the cable holder
b with extra margin.
Secure the cables in the cable holder b
giving the cable a certain slack so that it
does not apply the stress to the camera
connector portion.
} Explanations other than the camera connector
position are same as the case when combined
with DP22/DP27.
Camera connector position
DP22/DP27:
Back side
STM7-CU:
Right side
Front view of microscope (when combined with STM7-CU)
Front view of microscope (when combined with DP22/DP27)