93
STM7
Problem
Cause
Remedy
Page
l) Display shows "013F1511".
The X scale cable is disconnected or the
abnormality is detected in the scale.
Connect the X-scale cable properly
and press the X-RESET button to
cancel the error.
The counter data displayed after
the error is canceled is not a correct
value. Perform the measurement
again.
If you cannot solve the problem,
contact Olympus for assistance.
69
m) Display shows "013F1521".
The Y scale cable is disconnected or the
abnormality is detected in the scale.
Connect the Y-scale cable properly
and press the Y-RESET button to
cancel the error.
The counter data displayed after
the error is canceled is not a correct
value. Perform the measurement
again.
If you cannot solve the problem,
contact Olympus for assistance.
69
n) Display shows "013F1531".
The Z scale cable is disconnected or the
abnormality is detected in the scale.
Connect the Z-scale cable properly
and press the Z-RESET button to
cancel the error.
The counter data displayed after
the error is canceled is not a correct
value. Perform the measurement
again.
If you cannot solve the problem,
contact Olympus for assistance.
69
o) Display shows "013F0600".
The printer is in error. (out of paper, the
cover is open, etc.)
Clear the printer's error status.
–
9. Autofocus unit
a) Autofocusing cannot be
obtained.
(Desired point cannot be
focused.)
Specimen has low reflectance (3% or
less).
Focus on the specimen manually.
28
Specimen is not suitable for AF.
· Irregular specimen surface
· Thin, transparent structure
Focus on the specimen manually.
Or perform autofocusing on a less
irregular surface.
28,39
Objective other than specified is in use.
Use objectives specified as
autofocusing compatible.
(Refer to the instruction manual
provided with the autofocus unit.)
–
Dirt or stain on specimen.
Remove dirt, stain, etc.
12
b) Though autofocusing has
been completed normally,
specimen is not focused
when looked at visually.
Autofocusing position adjustment is
erroneous.
Adjust focusing position.
39
c) Autofocusing cannot be
switched START/STOP.
The connection cable may be unplugged. Check that the cable is connected
properly.