DMTA041-01EN [U8778479], Rev. C, May 2014
Chapter 5
42
The external DC power supply must be able to withstand up to 10 A of “in rush
current” for a few microseconds.
The 15 V to 24 V requirement is the voltage needed at the
DC
IN
connector site. The
length of the cable has an impact on the power supply’s nominal value. For example,
with a nominal value of 15 V and 30 meters of cable, the voltage at the connector site
could be lower than 14 V which is too low for the instrument to start up.
5.2
Operating Specifications
This section details the operating specifications of the TomoScan FOCUS LT.
5.2.1
Pulser/Receiver Differences and Limitations
The pulser/receiver specifications are given by model type. The main differences
between the TomoScan FOCUS LT models are the following:
•
Number of active channels (16, 32, or 64)
•
Channel quantity (32, 64, or 128)
AC power requirements
Voltage
100 VAC to 240 VAC, tolerance ±10 %
Frequency
50 Hz to 60 Hz, tolerance ±3 Hz
Maximum power
150 VA
Fuse
250 V, slow-blow:
1 A at 240 V
2 A at 120 V
DC power requirements
Voltage
15 VDC to 24 VDC,
tolerance
−
1 V, +1.5 V, 5 A maximum
Table 5 General specifications
(continued)
Category
Specification
Value
Summary of Contents for TomoScan FOCUS LT
Page 6: ...DMTA041 01EN U8778479 Rev C May 2014 Table of Contents vi...
Page 10: ...DMTA041 01EN U8778479 Rev C May 2014 Labels and Symbols 4...
Page 18: ...DMTA041 01EN U8778479 Rev C May 2014 Important Information Please Read Before Use 12...
Page 30: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 1 24...
Page 40: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 3 34...
Page 64: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 5 58...
Page 74: ...DMTA041 01EN U8778479 Rev C May 2014 List of Figures 68...
Page 76: ...DMTA041 01EN U8778479 Rev C May 2014 List of Tables 70...
Page 82: ...DMTA041 01EN U8778479 Rev C May 2014 Index 76...