DMTA041-01EN [U8778479], Rev. C, May 2014
Specifications
55
5.4
Specifications of the Ethernet Link
The TomoScan FOCUS LT must be linked to the workstation with a category 5e
Ethernet cable made of unshielded twisted pairs. The maximum cable length depends
on the link speed (see Table 12 on page 55).
Data type
A-scan type
RF, rectified, rectified video
C-scan type
4 gates (amplitude and time-of-flight for
gate peak and gate crossing C-scans), and a
synchronization gate (time-of-flight of the
echo)
Encoders
Maximum input frequency: 200 kHz
Interface modes: quadrature, clock
directions, up, down
Digital input
2 digital inputs (for more information, see
“I/O Connector” on page 60)
Digital output
4 digital outputs (for more information, see
“I/O Connector” on page 60)
Alarms
3 alarms available
Table 12 Ethernet cable parameters and specifications
Parameters
Specifications
Description
- Crossover Ethernet cable for direct
connection to the workstation
- Standard Ethernet cable for connection
to the workstation through a network hub
Olympus part number
Crossover cable: 60ND0001
Cable type
EIA/TIA standard, 150
Ω
, unshielded,
category 5e
Minimum distance between nodes
0.6 meters (2 feet)
Maximum cable length
100BASE-T: 100 meters (328 feet)
Table 11 Specifications of recorded data
(continued)
Category
Specification
Value
Summary of Contents for TomoScan FOCUS LT
Page 6: ...DMTA041 01EN U8778479 Rev C May 2014 Table of Contents vi...
Page 10: ...DMTA041 01EN U8778479 Rev C May 2014 Labels and Symbols 4...
Page 18: ...DMTA041 01EN U8778479 Rev C May 2014 Important Information Please Read Before Use 12...
Page 30: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 1 24...
Page 40: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 3 34...
Page 64: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 5 58...
Page 74: ...DMTA041 01EN U8778479 Rev C May 2014 List of Figures 68...
Page 76: ...DMTA041 01EN U8778479 Rev C May 2014 List of Tables 70...
Page 82: ...DMTA041 01EN U8778479 Rev C May 2014 Index 76...